Oxygen probe
CarboProbe HT

with ceramic sheath (%O2)

Description of the
oxygen probe
CarboProbe HT

Based on the ZrO2 C3M Electrolyte, the CarboProbe HT probe can handle temperatures up to 1700°C. All parts exposed to high temperatures are ceramic or platinum. This probe has a robust construction with a 15 mm diameter ceramic tube that protects the internal measurement device. This probe measures the oxygen concentration.

 

This probe cannot be used for the measurement of oxygen in carburising mixtures.

Maintenance

There are no maintenance parts required inside the sensor, if the platinum external connector is damaged it can be repaired without loss of signal quality.

Works great with our ControX display and control module

This probe is based on our ZrO2 mixed cell.

Technical

characteristics

Each of our probes is designed to meet specific needs, while offering the reliability and Swiss manufacturing quality characteristic of Econox.

Technical specifications

Output

0 to 1200 mv

Readout impedance

This probe should be used with controlling, recording and indicating instruments having input impedance of 8 megohms or higher

Accuracy

+- 2 mV in normal operating range

Response time

Less than 1 second

Thermocouple

R, S

Operating temperature

From 600°C (1100°F) to 1700°C (3100°F)

Mechanical shock

Resists mild mechanical shock. Handle carefully

Available lengths

300mm (11.8”), 400mm (15.7”), 500mm (19.7”), 600mm (23.6”), 700mm (27.5”), 800mm (31.5”), 900mm (35.4”), 1000mm (39.4), 1100mm (43.3”)

Reference air

Uncontaminated dry air at maximum rate of 1-6 l/h

External diameter

15 mm